13:08 - ECC is in state EccState.Ready with status EccStatus.NoError 13:08 - ECC temperature sensors: T1= 25.8, T2= 26.0, T3= 26.2, T4= 29.4 deg C 13:08 - 24 V power supply no. 1 is enabled with a current of 7.780 A. 13:08 - Checking high voltages... 13:08 - Starting task "Check HV" 13:08 - Checking reference voltages against 1500.000000 V 13:08 - Done 13:08 - Found 6 drawer(s). Reference voltage is between 1500.00 V and 1500.00 V. Actual voltage is between 1499.38 V and 1503.65 V. 13:08 - Maximum difference between reference and actual voltage is 3.04 V 13:08 - Checking status of HVPA module in 6 drawer(s)... 13:08 - Checking status of HVPA channels in 6 drawer(s)... 13:08 - HV is stabilized in all channels. 13:08 - Finished task "Check HV" 13:09 - Starting NMC 13:09 - Clearing PPS error counters 13:09 - Configuring TIB... 13:09 - Setting TIB forced mode to False 13:09 - Writing False to serveur.TIB.Forced.ForcedMode.ForcedMode_v 13:09 - Status: TibState.PpsStopped 13:09 - Alarms: AlarmVector.NoAlarm 13:09 - Writing 0.0 to serveur.TIB.ExternalDelay.ExternalDelay_v 13:09 - Setting TIB trigger mode to 0x8 = TriggerMask.SPhE 13:09 - Writing 8 to serveur.TIB.TriggerMode.ModeMask.ModeMask_v 13:09 - Writing 0 to serveur.TIB.InputsEnable.InputMask.InputMask_v 13:09 - Writing False to serveur.TIB.BUSY.DisableBusy.DisableBusy_v 13:09 - Writing 0.0 to serveur.TIB.BUSY.ExtraDelayBusy.ExtraDelayBusy_v 13:09 - Writing False to serveur.TIB.Throttler.DisableThrottlerBP.DisableThrottlerBP_v 13:09 - Writing 255 to serveur.TIB.Throttler.MaxTriggersBP50us.MaxTriggersBP50us_v 13:09 - Writing 255 to serveur.TIB.Throttler.MaxTriggersBP1ms.MaxTriggersBP1ms_v 13:09 - Writing False to serveur.TIB.Throttler.DisableThrottlerWR.DisableThrottlerWR_v 13:09 - Writing 255 to serveur.TIB.Throttler.MaxTriggersWR50us.MaxTriggersWR50us_v 13:09 - Writing 255 to serveur.TIB.Throttler.MaxTriggersWR1ms.MaxTriggersWR1ms_v 13:09 - Status: TibState.PpsStopped 13:09 - Alarms: AlarmVector.NoAlarm 13:09 - Configuring calibration box... 13:09 - Configuring calibration box with leds=0x40 voltage=12.00 frequency=1000.0/1 central_current=0.0 duration=500 width=1 lemo_out=True fiber1_out=True fiber2_out=True lemo_in=False enabled=False 13:09 - Closing TIB connection to camera server 13:09 - Connecting TIB to camera server at 10.10.0.251:222 13:09 - Starting TIB... 13:09 - Status: TibState.PpsStopped 13:09 - Alarms: AlarmVector.NoAlarm 13:09 - Resetting run 13:09 - Wait for UCTS external trigger before enabling trigger... 13:09 - Status: TibState.Standby 13:09 - Alarms: AlarmVector.NoAlarm 13:09 - Status: TibState.Standby 13:09 - Alarms: AlarmVector.NoAlarm 13:09 - Status: TibState.CoutingPps 13:09 - Alarms: AlarmVector.NoAlarm 13:09 - Starting TIB triggers... 13:09 - Enabling trigger 13:09 - Status: TibState.Running 13:09 - Alarms: AlarmVector.NoAlarm 13:09 - Starting flashes... 13:10 - Stopping TIB... 13:10 - Status: TibState.Running 13:10 - Alarms: AlarmVector.NoAlarm 13:10 - Disabling trigger 13:10 - Disabling PPS 13:10 - Status: TibState.PpsStopped 13:10 - Alarms: AlarmVector.NoAlarm 13:10 - Closing TIB connection to camera server 13:10 - Status: TibState.PpsStopped 13:10 - Alarms: AlarmVector.NoAlarm 13:10 - Run duration: 0:01:00 13:10 - Local camera triggers: 0 13:10 - Slow control triggers: 0 13:10 - Single photo-electron triggers: 49952 13:10 - Calibration triggers: 0 13:10 - Software triggers from UCTS: 0 13:10 - Auto-pedestal triggers: 389474 13:10 - Collected input triggers: 49952 13:10 - Busy triggers: 0 13:10 - Non-busy triggers sent to UCTS: 49952 13:10 - Camera event triggers sent to clusters: 49952 13:10 - Stopping NMC 13:10 - Found 0 PPS errors. 13:10 - Checking high voltages... 13:10 - Starting task "Check HV" 13:10 - Checking reference voltages against 1500.000000 V 13:10 - Done 13:10 - Found 6 drawer(s). Reference voltage is between 1500.00 V and 1500.00 V. Actual voltage is between 1499.38 V and 1503.65 V. 13:10 - Maximum difference between reference and actual voltage is 3.65 V 13:10 - Checking status of HVPA module in 6 drawer(s)... 13:10 - Checking status of HVPA channels in 6 drawer(s)... 13:10 - HV is stabilized in all channels. 13:10 - Finished task "Check HV"