17:00 - Checking high voltages... 17:00 - Starting task "Check HV" 17:00 - Checking reference voltages against /local/home/cta_test_bench/tests/2018/20180524/config_6modules.NMC.xml 17:00 - Done 17:00 - Found 6 drawer(s). Reference voltage is between 968.11 V and 1223.41 V. Actual voltage is between 968.68 V and 1224.27 V. 17:00 - Maximum difference between reference and actual voltage is 1.07 V 17:00 - Checking status of HVPA module in 6 drawer(s)... 17:00 - Checking status of HVPA channels in 6 drawer(s)... 17:00 - HV is stabilized in all channels. 17:00 - Finished task "Check HV" 17:00 - Rebooting TIB... 17:02 - Configuring TIB... 17:02 - Setting TIB forced mode to False 17:02 - Writing False to serveur.TIB.Forced.ForcedMode.ForcedMode_v 17:02 - Status: TibState.RunStopped 17:02 - Alarms: AlarmVector.NoAlarm 17:02 - Disabling PPS 17:02 - Status: TibState.PpsStopped 17:02 - Alarms: AlarmVector.NoAlarm 17:02 - Writing 0.0 to serveur.TIB.ExternalDelay.ExternalDelay_v 17:02 - Setting TIB trigger mode to 0x8 = TriggerMask.SPhE 17:02 - Writing 8 to serveur.TIB.TriggerMode.ModeMask.ModeMask_v 17:02 - Writing 0 to serveur.TIB.InputsEnable.InputMask.InputMask_v 17:02 - Writing False to serveur.TIB.BUSY.DisableBusy.DisableBusy_v 17:02 - Writing 0.0 to serveur.TIB.BUSY.ExtraDelayBusy.ExtraDelayBusy_v 17:02 - Writing False to serveur.TIB.Throttler.DisableThrottlerBP.DisableThrottlerBP_v 17:02 - Writing 255 to serveur.TIB.Throttler.MaxTriggersBP50us.MaxTriggersBP50us_v 17:02 - Writing 255 to serveur.TIB.Throttler.MaxTriggersBP1ms.MaxTriggersBP1ms_v 17:02 - Writing False to serveur.TIB.Throttler.DisableThrottlerWR.DisableThrottlerWR_v 17:02 - Writing 255 to serveur.TIB.Throttler.MaxTriggersWR50us.MaxTriggersWR50us_v 17:02 - Writing 255 to serveur.TIB.Throttler.MaxTriggersWR1ms.MaxTriggersWR1ms_v 17:02 - Status: TibState.PpsStopped 17:02 - Alarms: AlarmVector.NoAlarm 17:02 - Starting NMC 17:02 - Clearing PPS error counters 17:03 - Closing TIB connection to camera server 17:03 - Connecting TIB to camera server at 10.10.0.251:222 17:03 - Starting TIB... 17:03 - Status: TibState.PpsStopped 17:03 - Alarms: AlarmVector.NoAlarm 17:03 - Resetting run 17:03 - Wait for UCTS external trigger before enabling trigger... 17:03 - Status: TibState.Standby 17:03 - Alarms: AlarmVector.NoAlarm 17:03 - Status: TibState.Standby 17:03 - Alarms: AlarmVector.NoAlarm 17:03 - Status: TibState.CoutingPps 17:03 - Alarms: AlarmVector.NoAlarm 17:03 - Starting TIB triggers... 17:03 - Enabling trigger 17:03 - Status: TibState.Running 17:03 - Alarms: AlarmVector.NoAlarm 17:03 - Configuring calibration box... 17:03 - Configuring calibration box with leds=0x358 voltage=9.75 frequency=5000.0/1 central_current=0.0 duration=100 width=1 lemo_out=True fiber1_out=True fiber2_out=True lemo_in=False enabled=False 17:03 - Starting flashes... 17:03 - Stopping TIB... 17:03 - Status: TibState.Running 17:03 - Alarms: AlarmVector.NoAlarm 17:03 - Disabling trigger 17:03 - Disabling PPS 17:03 - Status: TibState.PpsStopped 17:03 - Alarms: AlarmVector.NoAlarm 17:03 - Closing TIB connection to camera server 17:04 - Status: TibState.PpsStopped 17:04 - Alarms: AlarmVector.NoAlarm 17:04 - Run duration: 0:00:49 17:04 - Local camera triggers: 0 17:04 - Slow control triggers: 0 17:04 - Single photo-electron triggers: 49709 17:04 - Calibration triggers: 0 17:04 - Software triggers from UCTS: 0 17:04 - Auto-pedestal triggers: 316505 17:04 - Collected input triggers: 49709 17:04 - Busy triggers: 0 17:04 - Non-busy triggers sent to UCTS: 49709 17:04 - Camera event triggers sent to clusters: 49709 17:04 - Stopping NMC 17:04 - Found 0 PPS errors. 17:04 - Checking high voltages... 17:04 - Starting task "Check HV" 17:04 - Checking reference voltages against /local/home/cta_test_bench/tests/2018/20180524/config_6modules.NMC.xml 17:04 - Done 17:04 - Found 6 drawer(s). Reference voltage is between 968.11 V and 1223.41 V. Actual voltage is between 968.68 V and 1224.27 V. 17:04 - Maximum difference between reference and actual voltage is 1.07 V 17:04 - Checking status of HVPA module in 6 drawer(s)... 17:04 - Checking status of HVPA channels in 6 drawer(s)... 17:04 - HV is stabilized in all channels. 17:04 - Finished task "Check HV"