14:00 - Starting task "Configure NMC" 14:00 - Configuring modules with configuration /local/home/cta_test_bench/tests/2018/20180523/config_11modules_SPE.NMC.xml 14:00 - NMC is in state NmcState.Configured and knows of 11 drawer(s) 14:00 - Found 0 PPS errors. 14:00 - Finished task "Configure NMC" 14:00 - Checking high voltages... 14:00 - Starting task "Check HV" 14:00 - Checking reference voltages against /local/home/cta_test_bench/tests/2018/20180523/config_11modules_SPE.NMC.xml 14:00 - Reference voltage of channel 3 in drawer #114: expected 1194.02, read 1196.38 14:00 - Reference voltage of channel 4 in drawer #131: expected 1138.15, read 1141.30 14:00 - Reference voltage of channel 0 in drawer #149: expected 1122.25, read 0.00 14:00 - Reference voltage of channel 4 in drawer #150: expected 1170.61, read 1172.04 14:00 - Done 14:00 - Found 11 drawer(s). Reference voltage is between 0.00 V and 1262.89 V. Actual voltage is between 0.00 V and 1264.53 V. 14:00 - Maximum difference between reference and actual voltage is 1.54 V 14:00 - Checking status of HVPA module in 11 drawer(s)... 14:00 - HVPA module in drawer 149 has a status of 0x7e! 14:00 - Checking status of HVPA channels in 11 drawer(s)... 14:00 - HVPA channel 0 of drawer 149 has status 0x8001! 14:00 - Resetting HVPA channel 0 of drawer 149... 14:00 - HV was locked in 1 channel(s), was in rise default in 0 channel(s) and was still in progress in 0 channel(s) 14:00 - Finished task "Check HV" 14:00 - Enabling high voltages... 14:00 - Starting task "Enable HV" 14:00 - Setting HV to 100.0 V 14:00 - Check of HVPA failed: expected 100.00, read 0.00 14:00 - Check of HVPA failed: expected 100.00, read 1064.30 14:00 - Check of HVPA failed: expected 100.00, read 1114.02 14:00 - Check of HVPA failed: expected 100.00, read 1238.24 14:00 - Check of HVPA failed: expected 100.00, read 1226.36 14:00 - Check of HVPA failed: expected 100.00, read 1226.36 14:00 - Checking status of HVPA channels in 11 drawer(s)... 14:00 - HVPA channel 0 of drawer 149 has status 0x8001! 14:00 - Resetting HVPA channel 0 of drawer 149... 14:00 - HV was locked in 1 channel(s), was in rise default in 0 channel(s) and was still in progress in 76 channel(s) 14:00 - Setting HV to nominal values... 14:00 - Setting channel 0 voltages to [1120.64, 1181.26, 1102.08, 1081.75, 1121.46, 1107.73, 1055.08, 1077.02, 1122.25, 1114.76, 1130.1] 14:00 - Setting channel 1 voltages to [1127.79, 1184.5, 1116.91, 1173.2, 1067.32, 1063.47, 1064.3, 1192.95, 1097.96, 1093.28, 1156.81] 14:00 - Setting channel 2 voltages to [1084.38, 1181.05, 1102.69, 1171.51, 1106.02, 1164.86, 1065.7, 1134.52, 1207.48, 1122.99, 1211.52] 14:00 - Setting channel 3 voltages to [1194.02, 1153.03, 1177.69, 1173.3, 1114.02, 1123.56, 1171.51, 1246.02, 1243.34, 1172.04, 1164.9] 14:00 - Setting channel 4 voltages to [1160.39, 1200.52, 1230.66, 1138.15, 1136.63, 1205.2, 1152.11, 1169.84, 1130.77, 1170.61, 1183.84] 14:00 - Setting channel 5 voltages to [1218.81, 1225.58, 1238.24, 1135.99, 1141.87, 1182.28, 1135.79, 1220.18, 1088.66, 1226.36, 1039.84] 14:00 - Setting channel 6 voltages to [1124.32, 1127.34, 1262.89, 1121.01, 1145.8, 1100.04, 1188.38, 1134.65, 1118.04, 1194.74, 1162.14] 14:00 - Checking status of HVPA channels in 11 drawer(s)... 14:00 - HVPA channel 0 of drawer 149 has status 0x8001! 14:00 - Resetting HVPA channel 0 of drawer 149... 14:01 - HV was locked in 1 channel(s), was in rise default in 0 channel(s) and was still in progress in 38 channel(s) 14:01 - Finished task "Enable HV" 14:01 - Checking high voltages... 14:01 - Starting task "Check HV" 14:01 - Checking reference voltages against /local/home/cta_test_bench/tests/2018/20180523/config_11modules_SPE.NMC.xml 14:01 - Done 14:01 - Found 11 drawer(s). Reference voltage is between 1039.84 V and 1262.89 V. Actual voltage is between 0.00 V and 1263.92 V. 14:01 - Maximum difference between reference and actual voltage is 1122.25 V 14:01 - Checking status of HVPA module in 11 drawer(s)... 14:01 - HVPA module in drawer 149 has a status of 0x7e! 14:01 - Checking status of HVPA channels in 11 drawer(s)... 14:01 - HVPA channel 0 of drawer 149 has status 0x8001! 14:01 - Resetting HVPA channel 0 of drawer 149... 14:01 - HV was locked in 1 channel(s), was in rise default in 0 channel(s) and was still in progress in 1 channel(s) 14:01 - Finished task "Check HV" 14:01 - Checking high voltages... 14:01 - Starting task "Check HV" 14:01 - Checking reference voltages against /local/home/cta_test_bench/tests/2018/20180523/config_11modules_SPE.NMC.xml 14:01 - Reference voltage of channel 0 in drawer #149: expected 1122.25, read 0.00 14:01 - Done 14:01 - Found 11 drawer(s). Reference voltage is between 0.00 V and 1262.89 V. Actual voltage is between 0.00 V and 1263.92 V. 14:01 - Maximum difference between reference and actual voltage is 14.55 V 14:01 - Checking status of HVPA module in 11 drawer(s)... 14:01 - Checking status of HVPA channels in 11 drawer(s)... 14:01 - HV was locked in 0 channel(s), was in rise default in 0 channel(s) and was still in progress in 1 channel(s) 14:01 - Finished task "Check HV" 14:01 - Checking high voltages... 14:01 - Starting task "Check HV" 14:01 - Checking reference voltages against /local/home/cta_test_bench/tests/2018/20180523/config_11modules_SPE.NMC.xml 14:01 - Reference voltage of channel 0 in drawer #149: expected 1122.25, read 0.00 14:01 - Done 14:02 - Found 11 drawer(s). Reference voltage is between 0.00 V and 1262.89 V. Actual voltage is between 0.00 V and 1263.31 V. 14:02 - Maximum difference between reference and actual voltage is 1.13 V 14:02 - Checking status of HVPA module in 11 drawer(s)... 14:02 - Checking status of HVPA channels in 11 drawer(s)... 14:02 - HV is stabilized in all channels. 14:02 - Finished task "Check HV" 14:02 - Starting NMC 14:02 - Clearing PPS error counters 14:03 - Configuring TIB... 14:03 - Setting TIB forced mode to False 14:03 - Writing False to serveur.TIB.Forced.ForcedMode.ForcedMode_v 14:03 - Status: TibState.PpsStopped 14:03 - Alarms: AlarmVector.NoAlarm 14:03 - Writing 0.0 to serveur.TIB.ExternalDelay.ExternalDelay_v 14:03 - Setting TIB trigger mode to 0x8 = TriggerMask.SPhE 14:03 - Writing 8 to serveur.TIB.TriggerMode.ModeMask.ModeMask_v 14:03 - Writing 0 to serveur.TIB.InputsEnable.InputMask.InputMask_v 14:03 - Writing False to serveur.TIB.BUSY.DisableBusy.DisableBusy_v 14:03 - Writing 0.0 to serveur.TIB.BUSY.ExtraDelayBusy.ExtraDelayBusy_v 14:03 - Writing False to serveur.TIB.Throttler.DisableThrottlerBP.DisableThrottlerBP_v 14:03 - Writing 255 to serveur.TIB.Throttler.MaxTriggersBP50us.MaxTriggersBP50us_v 14:03 - Writing 255 to serveur.TIB.Throttler.MaxTriggersBP1ms.MaxTriggersBP1ms_v 14:03 - Writing False to serveur.TIB.Throttler.DisableThrottlerWR.DisableThrottlerWR_v 14:03 - Writing 255 to serveur.TIB.Throttler.MaxTriggersWR50us.MaxTriggersWR50us_v 14:03 - Writing 255 to serveur.TIB.Throttler.MaxTriggersWR1ms.MaxTriggersWR1ms_v 14:03 - Status: TibState.PpsStopped 14:03 - Alarms: AlarmVector.NoAlarm 14:04 - Closing TIB connection to camera server 14:04 - Connecting TIB to camera server at 10.10.0.251:222 14:04 - Starting TIB... 14:04 - Status: TibState.PpsStopped 14:04 - Alarms: AlarmVector.NoAlarm 14:04 - Resetting run 14:04 - Wait for UCTS external trigger before enabling trigger... 14:04 - Status: TibState.Standby 14:04 - Alarms: AlarmVector.NoAlarm 14:04 - Status: TibState.Standby 14:04 - Alarms: AlarmVector.NoAlarm 14:04 - Status: TibState.Armed 14:04 - Alarms: AlarmVector.NoAlarm 14:04 - Status: TibState.CoutingPps ConnectionStatus: True 14:04 - Alarms: AlarmVector.NoAlarm 14:04 - Temperature: 36.25 14:04 - Local camera rate: 0.0 Hz 14:04 - Stereo trigger rate: 0.0 Hz 14:04 - SPE trigger rate: 0.0 Hz 14:04 - Calibration trigger rate: 0.0 Hz 14:04 - Software trigger rate: 0.0 Hz 14:04 - Auto-pedestal trigger rate: 6575.0 Hz 14:04 - Collected trigger rate: 0.0 Hz 14:04 - Busy rate: 0.0 Hz 14:04 - Camera trigger rate: 0.0 Hz 14:04 - Configuring calibration box... 14:04 - Configuring calibration box with leds=0x358 voltage=9.75 frequency=1000.0/1 central_current=0.0 duration=100 width=1 lemo_out=True fiber1_out=True fiber2_out=True lemo_in=False enabled=False 14:04 - Starting flashes... 14:04 - Status: TibState.CoutingPps ConnectionStatus: True 14:04 - Alarms: AlarmVector.NoAlarm 14:04 - Temperature: 36.19 14:04 - Local camera rate: 0.0 Hz 14:04 - Stereo trigger rate: 0.0 Hz 14:04 - SPE trigger rate: 71.0 Hz 14:04 - Calibration trigger rate: 0.0 Hz 14:04 - Software trigger rate: 0.0 Hz 14:04 - Auto-pedestal trigger rate: 6500.0 Hz 14:04 - Collected trigger rate: 0.0 Hz 14:04 - Busy rate: 0.0 Hz 14:04 - Camera trigger rate: 0.0 Hz 14:05 - Starting TIB triggers... 14:05 - Enabling trigger 14:05 - Status: TibState.Running 14:05 - Alarms: AlarmVector.NoAlarm 14:05 - Configuring calibration box... 14:05 - Configuring calibration box with leds=0x358 voltage=9.75 frequency=1000.0/1 central_current=0.0 duration=100 width=1 lemo_out=True fiber1_out=True fiber2_out=True lemo_in=False enabled=False 14:05 - Starting flashes... 14:05 - Configuring calibration box... 14:05 - Configuring calibration box with leds=0x358 voltage=9.75 frequency=1000.0/1 central_current=0.0 duration=100 width=1 lemo_out=True fiber1_out=True fiber2_out=True lemo_in=False enabled=False 14:05 - Starting flashes... 14:06 - Configuring calibration box... 14:06 - Configuring calibration box with leds=0x358 voltage=9.75 frequency=1000.0/1 central_current=0.0 duration=100 width=1 lemo_out=True fiber1_out=True fiber2_out=True lemo_in=False enabled=False 14:06 - Starting flashes... 14:06 - Configuring calibration box... 14:06 - Configuring calibration box with leds=0x358 voltage=9.75 frequency=1000.0/1 central_current=0.0 duration=100 width=1 lemo_out=True fiber1_out=True fiber2_out=True lemo_in=False enabled=False 14:06 - Starting flashes... 14:06 - Configuring calibration box... 14:06 - Configuring calibration box with leds=0x358 voltage=9.75 frequency=1000.0/1 central_current=0.0 duration=100 width=1 lemo_out=True fiber1_out=True fiber2_out=True lemo_in=False enabled=False 14:06 - Starting flashes... 14:07 - Stopping TIB... 14:07 - Status: TibState.Running 14:07 - Alarms: AlarmVector.NoAlarm 14:07 - Disabling trigger 14:07 - Disabling PPS 14:07 - Status: TibState.PpsStopped 14:07 - Alarms: AlarmVector.NoAlarm 14:07 - Closing TIB connection to camera server 14:07 - Status: TibState.PpsStopped 14:07 - Alarms: AlarmVector.NoAlarm 14:07 - Run duration: 0:02:57 14:07 - Local camera triggers: 0 14:07 - Slow control triggers: 0 14:07 - Single photo-electron triggers: 59743 14:07 - Calibration triggers: 0 14:07 - Software triggers from UCTS: 0 14:07 - Auto-pedestal triggers: 1147559 14:07 - Collected input triggers: 49797 14:07 - Busy triggers: 0 14:07 - Non-busy triggers sent to UCTS: 49797 14:07 - Camera event triggers sent to clusters: 49797 14:07 - Stopping NMC 14:07 - Found 0 PPS errors. 14:07 - Checking high voltages... 14:07 - Starting task "Check HV" 14:07 - Checking reference voltages against /local/home/cta_test_bench/tests/2018/20180523/config_11modules_SPE.NMC.xml 14:07 - Reference voltage of channel 0 in drawer #149: expected 1122.25, read 0.00 14:07 - Done 14:07 - Found 11 drawer(s). Reference voltage is between 0.00 V and 1262.89 V. Actual voltage is between 0.00 V and 1263.92 V. 14:07 - Maximum difference between reference and actual voltage is 1.13 V 14:07 - Checking status of HVPA module in 11 drawer(s)... 14:07 - Checking status of HVPA channels in 11 drawer(s)... 14:07 - HV is stabilized in all channels. 14:07 - Finished task "Check HV"